Verification procedures, purpose and repeat criteria for verifying BpTRU™,16 Canadian Health Measures Survey, cycle 1, 2007 to 2009

Standard table symbols

Verification procedure Purpose Repeat criteria
Zero calibration check Verifies that BpTRU™ accurately measures zero pressure when no pressure is being applied If value in PULSE display not equal to 0 or if value in READING display >10.
Reference gauge calibration check Compares accuracy of BpTRU™’s pressure transducer against reference pressure gauge (for example, mercury sphygmomanometer) of known accuracy when pressure applied to device Repeat based on difference between expected and actual readings at each pressure tested as follows:
275 > ± 5.5 mmHg
200 > ± 4 mmHg
50, 100 and 150 > ± 3 mmHg
Over pressure test Verifies BpTRU™’s safety mechanism that automatically releases air in cuff and displays error code should pressure surpass 330 mmHg Error code E2 should be displayed in SYSTOLIC display window; if not, repeat test.
Over infiation time test Verifies BpTRU™’s safety mechanism that automatically releases air in cuff if it stays pressurized over 10 mmHg for more than 180 seconds Error code E11 should be displayed in SYSTOLIC display window; if not, repeat test.